Evaluation of a Viscoelastic Coagulation Monitoring System (VCM Vet(®)) and Its Correlation with Thromboelastometry (ROTEM(®)) in Diseased and Healthy Dogs

对粘弹性凝血监测系统(VCM Vet®)及其与血栓弹力图(ROTEM®)在患病犬和健康犬中的相关性进行评估

阅读:1

Abstract

Thromboelastometry provides a real-time assessment of global coagulation in whole blood. A novel bed-side viscoelastic coagulation monitor (VCM) has been developed for use in small animals. The aims of the study were to determine inter-device agreement of two VCM devices, to evaluate the correlation between VCM and rotational thromboelastometry as a reference method (ROTEM), and to determine the sensitivity and specificity of VCM to diagnose hypo-, normo-, and hypercoagulability. ROTEM (extrinsic and intrinsic activation) analysis was performed using citrated blood and VCM analysis using native blood. Twenty healthy and forty diseased dogs with and without coagulopathies were enrolled. The VCM inter-device agreement was moderate to strong for most of the parameters, depending on the grading scale. Correlation between VCM and ROTEM was moderate to strong for parameters of clotting time and clot strength. The VCM most likely detects true hypocoagulability and reliably rules out hypocoagulability. The VCM has a high sensitivity in diagnosing normocoagulability, but incorrectly classified dogs with abnormal coagulation as normocoagulable. The VCM was not able to detect hypercoagulability. ROTEM and VCM cannot be used interchangeably.