Metallosis-Induced Warm Antibody Auto-Immune Hemolytic Anemia After Bilateral, Large-Diameter Metal-on-Metal Total Hip Arthroplasty With Complete Remission After Revision

双侧大直径金属对金属全髋关节置换术后金属中毒诱发的温抗体自身免疫性溶血性贫血,翻修术后完全缓解

阅读:1

Abstract

The use of metal-on-metal bearing couples in total hip arthroplasty can lead to an increased release of metal ions, particularly cobalt and chromium over time. This can lead to local and systemic metallosis, which has cytotoxic, genotoxic, and immunotoxic effects and can cause a host of secondary disorders. We describe the case of a 37-year-old female patient that was diagnosed with warm-antibody autoimmune hemolytic anemia (WAIHA) one and a half years after bilateral large-diameter head metal-on-metal total hip arthroplasty. For 11 years, it was refractory to all therapy, including splenectomy and rituximab, requiring long-term oral prednisone for disease control. Ultimately, systemic metallosis and periprosthetic joint infection were diagnosed, requiring explantation of the prostheses. By the sixth week postoperatively, she experienced complete spontaneous remission of her WAIHA. In conclusion, WAIHA can be associated with systemic metallosis in patients with metal-on-metal prosthetic joint replacements. Both hematologists and orthopedic surgeons should be aware of this.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。