Evaluation of the Abbott Alinity i Thyroid-Stimulating Hormone Receptor Antibody (TRAb) Chemiluminescent Microparticle Immunoassay (CMIA)

雅培 Alinity i 促甲状腺激素受体抗体 (TRAb) 化学发光微粒免疫测定 (CMIA) 的评价

阅读:2

Abstract

Background: We evaluated the performance of the Abbott thyroid-stimulating hormone receptor antibody chemiluminescent microparticle immunoassay (CMIA) on the Alinity i. Methods: Verification studies for precision, linearity, analytical measuring range, diagnostic cut offs for Graves' disease were performed. We compared the Abbott CMIA to an established TRAb assay (Roche electrochemiluminescence immunoassay). Method comparison analysis was performed between serum and plasma samples on the Abbott CMIA. Results: Repeatability (CV%) for TRAb were 4.07, 1.56, 0.71 and within-laboratory imprecision (CV%) were 4.07, 1.90, 0.71 at 3.0, 10.0, 30.0 IU/L of TRAb, respectively. Linearity and analytical measuring range were verified from 1.07-47.9 IU/L. The limit of the blank was 0 IU/L, limit of detection was 0.15 IU/L, and limit of quantification was 0.5 IU/L. Passing-Bablok analysis showed agreement between the two assays; Y-intercept = 0.787, slope = 1.04. Passing-Bablok analysis also showed agreement between the plasma and serum samples run on the Abbott CMIA; Y-intercept -0.17, slope = 0.97. Conclusions: The Abbott TRAb CMIA on the Alinity i performs within the manufacturer claims for assay precision, linearity, analytical measuring range, limit of blank, limit of detection, limit of quantitation and diagnostic cut offs for Graves' disease. Thus, the Abbott TRAb CMIA on the Alinity i is fit for clinical use.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。