A device for surveillance of vascular access sites for bleeding: results from a clinical evaluation trial

用于监测血管通路部位出血的装置:一项临床评估试验的结果

阅读:2

Abstract

Post-procedural wound haemorrhage is a potentially life-threatening complication. For haemodialysis patients, bleeding is often encountered after vascular access procedures and fatal episodes have been reported. Visual monitoring for bleeding is manpower intensive and bleeding episodes may still be missed between inspections. A device, Blood WArning Technology with Continuous Haemoglobin sensor (BWATCH), was developed to detect bleeding from wounds. This a prospective, observational clinical trial on patients who have had a dialysis catheter inserted or removed. The battery-powered, disc-shaped device (43 mm diameter, 12 mm height) was placed over the dressing for at least six hours. The device detects reflected light with characteristics specific for haemoglobin and an alarm would be triggered if bleeding occurs. There were 250 participants (177 post-insertion, 73 post-removal) and 36 episodes of bleeding occurred. The device alarm was triggered in all instances but there were also 9 false alarms. Specificity was 95.8%, false positive rate was 4.2% and positive predictive value was 80.0%. Sensitivity and negative predictive value were 100% but detection failure may still occur due to improper application or device maintenance. The use of technological aids for monitoring improves patient safety and may reduce demand on manpower.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。