Contamination of the central medical air supply with water leading to mass ventilator failure

中央医用空气供应系统被水污染,导致大量呼吸机故障

阅读:1

Abstract

Here, we present a case of mass ventilator failure due to contaminated medical air. Multiple ventilators failed routine tests, including almost all of the ventilators in our intensive care unit. A faulty air compressor had led to water contamination of our centre's supply of medical air. Water entered the pipeline supply of air and, hence the ventilators and anaesthetic machines. The disruption of the machines' proportional mixer valve resulted in unreliable delivery of fresh gas flow. This malfunction was discovered during routine pre-use checks, and backup ventilators were available to replace the faulty ventilators. A shortage of equipment was averted due to a serendipitous availability of ventilator stockpiles prepared for the COVID-19 pandemic. Ventilator shortages are commonly described in mass casualty and pandemic scenarios. While there are multiple strategies described in literature to augment and maximise equipment available for mechanical ventilation, stockpiling equipment remains an expensive but necessary component of disaster contingency planning.