The double-lasso loop technique of Biceps tenodesis has lower displacement after cyclic loading, compared to interference screw fixation: Biomechanical analysis in an ovine model

与干涉螺钉固定相比,双套索环固定法在循环载荷后肱二头肌腱固定术的位移更小:绵羊模型生物力学分析

阅读:1

Abstract

BACKGROUND: Biceps tenodesis is an effective treatment for symptomatic long head of biceps tendon pathology. Recently the arthroscopic "double lasso-loop" suture anchor (DLSA) technique was described, advantaged by reduced cost, complexity, and operative time. We aimed to compare the in vitro strength of DLSA with conventional interference screws (IS). METHODS: A biomechanical analysis was conducted on 14 sheep shoulders (8 DLSA, 6 IS), consisting of a 500-cycle cyclic loading experiment of 5-70 N and ultimate failure load (UFL) test where each specimen was pulled until failure. Displacement (mm) was recorded every 100 cycles, while stiffness and UFL were observed. RESULTS: Cyclic displacement was significantly lower with DLSA at 100 cycles, but not above. During the UFL test, IS was stiffer (27.68 ± 6.56 N/mm versus 14.10 ± 5.80 N/mm, p = .005) and had higher UFL (453.67 ± 148.55 N versus 234.22 ± 44.57 N, p = .001) than DLSA. All DLSA failures occurred with suture/anchor pull-out, while all IS constructs failed at the muscle/tendon. DISCUSSION: Comparison of the novel DLSA technique with a traditional IS method found lower initial displacement. While our IS constructs could withstand higher UFL, peak load characteristics of DLSA were similar to previous ovine studies. Hence, the DLSA technique is a viable alternative to IS for biceps tenodesis with its purported non-biomechanical advantages.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。