Accuracy of an artificial intelligence algorithm for detecting moderate-to-severe vertebral compression fractures on abdominal and thoracic computed tomography scans

人工智能算法在腹部和胸部计算机断层扫描中检测中重度椎体压缩性骨折的准确性

阅读:1

Abstract

OBJECTIVE: To describe the accuracy of HealthVCF, a software product that uses artificial intelligence, in the detection of incidental moderate-to-severe vertebral compression fractures (VCFs) on chest and abdominal computed tomography scans. MATERIALS AND METHODS: We included a consecutive sample of 899 chest and abdominal computed tomography scans of patients 51-99 years of age. Scans were retrospectively evaluated by the software and by two specialists in musculoskeletal imaging for the presence of VCFs with vertebral body height loss > 25%. We compared the software analysis with that of a general radiologist, using the evaluation of the two specialists as the reference. RESULTS: The software showed a diagnostic accuracy of 89.6% (95% CI: 87.4-91.5%) for moderate-to-severe VCFs, with a sensitivity of 73.8%, a specificity of 92.7%, and a negative predictive value of 94.8%. Among the 145 positive scans detected by the software, the general radiologist failed to report the fractures in 62 (42.8%), and the algorithm detected additional fractures in 38 of those scans. CONCLUSION: The software has good accuracy for the detection of moderate-to-severe VCFs, with high specificity, and can increase the opportunistic detection rate of VCFs by radiologists who do not specialize in musculoskeletal imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。