Monte Carlo derivation of filtered tungsten anode X-ray spectra for dose computation in digital mammography

利用蒙特卡罗方法推导滤波钨阳极X射线光谱,用于数字乳腺摄影剂量计算

阅读:1

Abstract

OBJECTIVE: Derive filtered tungsten X-ray spectra used in digital mammography systems by means of Monte Carlo simulations. MATERIALS AND METHODS: Filtered spectra for rhodium filter were obtained for tube potentials between 26 and 32 kV. The half-value layer (HVL) of simulated filtered spectra were compared with those obtained experimentally with a solid state detector Unfors model 8202031-H Xi R/F & MAM Detector Platinum and 8201023-C Xi Base unit Platinum Plus w mAs in a Hologic Selenia Dimensions system using a direct radiography mode. RESULTS: Calculated HVL values showed good agreement as compared with those obtained experimentally. The greatest relative difference between the Monte Carlo calculated HVL values and experimental HVL values was 4%. CONCLUSION: The results show that the filtered tungsten anode X-ray spectra and the EGSnrc Monte Carlo code can be used for mean glandular dose determination in mammography.