Bioresorbable scaffold -fourth revolution or failed revolution: Is low scaffold strut thickness the wrong target?

生物可吸收支架——第四次革命还是失败的革命:低支架支柱厚度是否是错误的目标?

阅读:1

Abstract

Bioresorbable scaffold (BRS) technology has currently fallen into disrepute because of inordinately high risk of scaffold thrombosis and post-procedure myocardial infarction. Low tensile and radial strengths of polymeric BRS contributing to improper strut embedment have been identified as major correlates of poor outcomes following BRS implantation. Magnesium has a better tensile/radial strength compared with polymeric BRS but it is still far lower than cobalt-chromium. Newers innovations utilizing alteration in polymer composition and orientation or even newer polymers have focused on attempts to reduce strut thickness but may have little effect on tensile/radial strength of finished product and therefore may not impact the BRS outcome on long run. Currently, newer generation BRS usage may be restricted to suitable low risk younger patients with proper vessel preparation and application of technique.