Management and Outcomes of Breakage in Polyethylene-Ceramic Composite (Sandwich-type) Liners: Lessons From Long-term Follow-up

聚乙烯-陶瓷复合材料(三明治型)衬垫破损的管理和结果:长期随访的经验教训

阅读:1

Abstract

This case series reports the only 3 instances of ceramic liner breakage among patients who underwent total hip arthroplasty (THA) with polyethylene-ceramic composite (sandwich-type) liners at our center between 1999 and 2002. Breakages occurred at approximately 6 (n = 2) and 17 years (n = 1) after the primary THA. Symptoms included audible crepitation, pain, and restricted motion during normal activities. Surgical revisions utilized a polyethylene liner and third-generation alumina ceramic head, with synovectomy and irrigation to remove debris (preventing third-body wear). All patients achieved good long-term outcomes. Although infrequent, the risk of breakage increases over time, and several patients still carry these liners years after THA. This case series underscores the importance of vigilant follow-up, patient education, and timely intervention to manage this infrequent but potentially catastrophic complication.