Accuracy of Leg Length and Hip Offset Measurements Using a Fluoroscopic Grid During Anterior Approach Total Hip Arthroplasty

前路全髋关节置换术中采用透视网格测量腿长和髋关节偏移的准确性

阅读:1

Abstract

BACKGROUND: Minimizing leg length (LLD) and hip offset (OD) discrepancies is critical for tissue tension and implant longevity in total hip arthroplasty (THA). The direct anterior approach (DAA) helps surgeons recreate these values under fluoroscopy. Several methods to accomplish this have been described, with no consensus on which is superior. This study evaluated the ability to minimize LLD and OD using a surgeon-controlled, adjustable fluoroscopic grid. We hypothesized that this tool would recreate parameters to within 10 mm of the contralateral side. METHODS: One hundred eleven primary THAs performed with an adjustable radiopaque grid to equalize leg length and hip offset were retrospectively reviewed. These values were measured on postoperative radiographs and compared to the contralateral hip. Patients were excluded if they had inadequate imaging, revision arthroplasty, preexisting deformities, or underwent approaches other than DAA. RESULTS: Mean age was 59.1 ± 11.1 years, 63.1% of patients were female, and average body mass index was 27.8 ± 7.0. Mean LLD was 3.7 ± 3.0 mm, while mean OD was 4.6 ± 3.6 mm. 95.5% of hips showed LLD < 10 mm, while 93.7% of hips had OD < 10 mm. Furthermore, 76.6% of hips had LLD < 5 mm, while 62.2% of hips had OD < 5 mm. CONCLUSIONS: The described technique restored limb length and hip offset during DAA THA. This technique yields consistent results and offers an inexpensive alternative to costly digital software and more cumbersome fixed grid systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。