An Atypical Adverse Local Tissue Reaction Seen With Metal-On-Metal Total Hip Arthroplasty Utilizing Polyethylene Liners With Metal Inlays

使用金属嵌体聚乙烯衬垫进行金属对金属全髋关节置换术时出现的一种非典型局部组织不良反应

阅读:1

Abstract

Adverse local tissue reaction (ALTR) is a frequently described, although uncommon, complication of metal-on-metal total hip arthroplasty. Here in we report on 2 patients with unique metal-inlay polyethylene liners who suffered from ALTR that required revision arthroplasty. In 2 of 3 cases the femoral trunnion was noted to have minimal corrosion and the stem was salvaged with a titanium adapter. In one case there was catastrophic femoral stem failure and an extended trochanteric osteotomy was required to remove the unique lateral flare stem. The surgeon must pay special attention when scrutinizing radiographs to identify a metal inlay polyethylene liner and when performing revision arthroplasty to prevent greater trochanter fracture in a patient who likely already has compromised abductor function due to ALTR.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。