The effect of ablation settings on lesion characteristics with DiamondTemp ablation system: An ex vivo experiment

DiamondTemp消融系统消融参数对病灶特征的影响:一项离体实验

阅读:1

Abstract

INTRODUCTION: Creating large lesion in ablations using the DiamondTemp (DTA) ablation system may reduce the frequency of arrhythmia recurrence and allow the treatment of ventricular arrhythmias. Therefore, this study aimed to investigate whether power, application time, contact force (CF), and contact angle affect lesion formation in the ventricles. METHODS: Ablations were delivered to porcine myocardial preps to evaluate the lesion characteristics. Ablations were conducted with a maximum power of 50 W, target temperature of 58°C, CF of 10, 20, or 30 g, and contact angle between the catheter tip and tissue. The ablation durations were 15, 30, 60 s, 15 s × 2, or 30 s × 2. RESULTS: Steam pops occurred only in cases with perpendicular contact. The lesion depth was larger in all settings in the perpendicular orientation than in the parallel orientation. The temperatures were lower in all settings in the perpendicular orientation than in the parallel orientation. The lesions became larger as CF increased with perpendicular contact and duration of ≥30 s. The longer application time resulted in larger surface area, depth, and volume of the lesion. Lesion depth was greater with single application of 30 and 60 s than with 15 s × 2 and 30 s × 2, respectively. CONCLUSION: It is important to perform a single prolonged application as much as possible to create deeper lesions. Parallel contact with the tissue should be maintained to take advantage of the temperature sensor's capabilities to avoid pop phenomenon.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。