Difference in radiofrequency ablation profile between during sinus rhythm and atrial fibrillation: Considerations in this era of high-power short-duration strategy

窦性心律和房颤时射频消融曲线的差异:高功率短时程策略时代的考量

阅读:2

Abstract

BACKGROUND: The concept of ablation index (AI) was introduced to evaluate radiofrequency (RF) ablation lesions. It is calculated from power, contact force (CF), and RF duration. However, other factors may also affect the quality of ablation lesions. To examine the difference in RF lesions made during sinus rhythm (SR) and atrial fibrillation (AF). METHODS: Sixty patients underwent index pulmonary vein isolation during SR (n = 30, SR group) or AF (n = 30, AF group). All ablations were performed with a power of 50 W, a targeted CF of 5-15 g, and AI of 400-450 using Thermocool Smarttouch SF. The CF, AI, RF duration, temperature rise (Δtemp), impedance drop (Δimp), and the CF stability of each ablation point quantified as the standard deviation of the CF (CF-SD) were compared between the two groups. RESULTS: A total of 3579 ablation points were analyzed, which included 1618 and 1961 points in the SR and the AF groups, respectively. Power, average CF, RF duration per point, and the resultant AI (389 ± 59 vs. 388 ± 57) were similar for the two rhythms. However, differences were seen in the CF-SD (3.5 ± 2.2 vs. 3.8 ± 2.1 g, p < .01), Δtemp (3.8 ± 1.3 vs. 4.0 ± 1.3°C, p < .005), and Δimp (10.3 ± 5.8 vs. 9.4 ± 5.4 Ω, p < .005). CONCLUSIONS: Despite similar AI, various RF parameters differed according to the underlying atrial rhythm. Ablation delivered during SR demonstrated less CF variability and temperature increase and greater impedance drop than during AF.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。