Low fluoroscopy permanent His bundle pacing using a new electroanatomic mapping system (KODEX EPD). A multicenter experience

使用新型电解剖标测系统(KODEX EPD)进行低透视永久性希氏束起搏:一项多中心经验

阅读:2

Abstract

BACKGROUND: His bundle pacing (HBP) may be a challenging procedure, often involving a long fluoroscopic time (FT) and a long procedural time (PT). We sought to evaluate whether the use of a new nonfluroscopic mapping (NFM) system, the KODEX-EPD, is able to reduce FT and PT when mapping is performed by the pacing catheter rather than an electrophysiological mapping catheter. METHODS AND RESULTS: We included 46 consecutive patients (77 ± 8 years; 63% male) who underwent HBP; in 22 a NFM-guided procedure with the KODEX-EPD system was performed (group 1), whereas in 24 a conventional fluoroscopy-guided approach was used (group 2). Pacing indications were sick sinus syndrome in 13, atrioventricular block in 21, and cardiac resynchronization therapy in 12 cases. Both a lumen-less fixed helix lead and a stylet-driven extendable helix lead were used, respectively, in 24% and 76% of patients. HBP was successful in 22 patients (100%) in group 1 and 23 patients (96%) in group 2. The FT was significantly reduced in group 1 (183 ± 117 s vs 464.1 ± 352 s in group 2, p = .012). There were no significant differences between groups in PT and other procedural outcomes. CONCLUSIONS: The KODEX-EPD system may be safely used in HBP procedures. It is effective in reducing ionizing radiation exposure, as evidenced by the significant drop in FT, without increasing PT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。