Electrocardiographic characteristics for the prediction of under-sensing in implantable loop recorders

植入式循环记录仪低感知预测的心电图特征

阅读:1

Abstract

BACKGROUND: Under-sensing (US) in implantable loop recorders (ILRs) interferes with the accurate diagnosis of arrhythmia, but there is little information available on the details of US of ILRs. The aim of this study was to clarify the frequency of US in patients with ILRs and to investigate the predictors of US in ILRs prior to implantation. METHODS AND RESULTS: We studied 46 consecutive patients implanted ILR. During the mean follow-up period of 499 ± 363 days, 15 events of US were observed in five patients. There were no significant differences in patient characteristics between patients with and without US. In standard 12-lead electrocardiogram (ECG), QRS complex amplitude in anterolateral chest leads (V2 to V5) were significantly lower in patients with than without US (V2: 0.88 [0.66, 1.22] mV vs. 1.67 [1.23, 2.29] mV, p = .010 V3: 1.25 [1.00, 1.26] mV vs. 1.90 [1.41, 2.29] mV, p = .013; V4: 1.14 [0.96, 1.38] mV vs. 1.93 [1.65, 2.64] mV, p = .023; V5: 0.57 [0.50, 0.75] mV vs. 1.60 [1.20, 1.98] mV, p = .011, respectively). ROC curve analysis showed that cut-off values of 1.30 mV of QRS complex amplitude in V2, 1.26 mV of that in V3, and 0.75 mV of that in V5 had moderate accuracy for predicting US (V2: sensitivity 68%, specificity 100%, area under the curve [AUC] 0.86; V3: sensitivity 85%, specificity 80%, AUC 0.85; V5: sensitivity 98%, specificity 80%, AUC 0.85, respectively). CONCLUSIONS: US was observed in 10.9% patients with an ILR. QRS complex amplitude in anterolateral chest leads (V2 to V5) on ECG might be useful for predicting US in patients with ILRs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。