Leadless pacemaker through tricuspid bioprosthetic valve: Early experience

经三尖瓣生物瓣膜植入无导线起搏器:早期经验

阅读:1

Abstract

BACKGROUND: Leadless pacemaker (LP) therapy has been proved effective in cases where traditional transvenous right pacing (TRP) failed. TRP through a bioprosthetic tricuspid valve (BTV) has always been considered an unpreferable solution because of possible deleterious effect of permanent pacing leads on BTV function and specifically on tricuspid regurgitation (TR). Very limited data exist about the feasibility and safety of LP implantation in this setting. METHODS: We describe two cases of LP implantation through BTV in patients with failure of epicardial pacemaker implanted after cardiac surgery. The focus is on technical description of the procedure and on electrical and echocardiographic evaluation at implantation and at the follow-up. RESULTS: In both cases, skilled and careful handling of the delivery system as well as proper use of X-ray oblique views was determinant for atraumatic successful valve crossing. Likewise, an accurate selection of the deployment site inside the right ventricle, far enough from the valve to avoid valvular dysfunction, was important for successful implantation. Electrical parameters of LP were satisfying at implantation and at the follow-up. The echocardiogram after implantation and at the follow-up showed no mechanical interference of LP with prosthetic valve, no significant TR, and absence of significant changes in the biventricular function. CONCLUSION: Our data seem to support feasibility and safety of this type of procedure in skilled hands, allowing efficacious pacing without valvular dysfunction or right ventricular (RV) physiology impairment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。