Investigation of optimal position for implantable loop recorders by potential mapping with Reveal DX

利用 Reveal DX 进行电位映射,研究植入式循环记录仪的最佳位置

阅读:1

Abstract

BACKGROUND: Electrode positioning is important for the operation of implantable loop recorders (ILRs). This study aimed to investigate optimal electrode positions for the implantation of ILRs. METHODS: Fifteen male subjects participated in the study and R wave amplitude data were collected from 15 areas in the left anterior chest area, with 4 variations of electrode angles and body positions. RESULTS: The estimated minimum R wave amplitude values were greater than 0.3 mV (manufacturer-recommended value) when electrodes were angled vertically and positioned on the midclavicular line of the third and the forth intercostal spaces or on the left sternal border of the fourth and fifth intercostal space and when angled at 45° to the right on the left sternal border of the third and the forth intercostal space. CONCLUSIONS: Exploring areas around the left sternal border of the fourth intercostal space - where the R wave amplitudes are least affected by body positions - is a reliable method for choosing implantation locations for ILRs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。