Quality specifications and their daily application to evaluate the accuracy of reference measurements for serum concentrations of 25-hydroxyvitamin D(3) and 25-hydroxyvitamin D(2)

质量规范及其在日常应用中对血清中 25-羟基维生素 D(3) 和 25-羟基维生素 D(2) 浓度参考测量准确性的评价

阅读:1

Abstract

BACKGROUND: Reference measurement procedures (RMP) have rigorous accuracy specifications. For total 25-hydroxyvitamin D, 25(OH)D, bias ≤1.7% and CV ≤5% are recommended. These quality specifications are impractical for minor analytes, such as 25(OH)D(2). Furthermore, documentation on RMP quality performance specifications for the individual 25(OH)D metabolites and their daily application are missing. METHODS: To assess accuracy, we used zeta-scores. Daily, 5-10 specimens (duplicate) and 3 reference materials (singleton or duplicate) were measured for 25(OH)D(3) and 25(OH)D(2) using JCTLM-accepted LC-MS/MS RMPs. Protocols were repeated on 3-4 occasions to generate campaign results. We used separate zeta-score acceptability criteria for daily (≤|2|) and campaign (≤|1|) evaluations. Allowable imprecision was determined experimentally. RESULTS: Across 7 campaigns, unacceptable daily zeta-scores required repeating 2 runs for 25(OH)D(3) and 5 runs for 25(OH)D(2). Hence, the zeta-scores of acceptable reference material results indicated high accuracy. The allowable imprecision for the RMPs was ≤5% (daily) and ≤ 3% (campaign) for 25(OH)D(3) and ≤ 7% (daily) and ≤ 4% (campaign) for 25(OH)D(2), respectively. CONCLUSIONS: Using zeta-scores and experimentally derived imprecision, we developed a straightforward approach to assess the acceptability of individual 25(OH)D reference measurements, providing also much-needed practical accuracy specifications for 25(OH)D(2).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。