Development of a rapid and sensitive LC-MS/MS assay for the determination of sorafenib in human plasma

开发一种快速灵敏的 LC-MS/MS 检测方法,用于测定人血浆中的索拉非尼

阅读:4
作者:Lokesh Jain, Erin R Gardner, Jürgen Venitz, William Dahut, William D Figg

Abstract

A rapid and sensitive liquid chromatography/tandem mass spectrometric (LC/MS/MS) assay was developed for the quantitative determination of sorafenib in human plasma. Sample pretreatment involved simple protein precipitation by the addition of 0.5 mL acetonitrile, containing internal standard ([2H3, 15N] sorafenib), to 50 microL of plasma sample volume. Separation was achieved on a Waters SymmetryShield RP8 (2.1 mm x 50 mm, 3.5 microm) column at room temperature using an isocratic elution method with acetonitrile/0.1% formic acid in water: 65/35 (v/v) at a flow rate of 0.25 mL/min. Detection was performed using electrospray ionization in positive ion multiple reaction monitoring (MRM) mode by monitoring the ion transitions from m/z 464.9 --> 252.0 (sorafenib) and m/z 469.0 --> 259.0 (internal standard). Calibration curves were linear in the concentration range of 5-2000 ng/mL. The accuracy and precision values, calculated from three different sets of quality control samples analyzed in quintuplicate on six different days, ranged from 92.86% to 99.88% and from 1.19% to 4.53%, respectively.