Targeted metabolic profiling of urinary steroids with a focus on analytical accuracy and sample stability

针对尿液类固醇的代谢分析,重点关注分析准确性和样品稳定性

阅读:9
作者:Nora Vogg, Tobias Müller, Andreas Floren, Thomas Dandekar, Oliver Scherf-Clavel, Martin Fassnacht, Matthias Kroiss, Max Kurlbaum

Conclusions

A sensitive and robust LC-MS/MS method for the quantification of 11 urinary steroids was developed and validated according to international guidelines. Pre-analytical matrix stability was evaluated and the suitability of the method for the analysis of clinical samples and prospective validation studies was shown.

Methods

A method for the quantification of 11 deconjugated steroids (5-pregnenetriol, dehydroepiandrosterone, cortisone, cortisol, α-cortolone, tetrahydro-11-deoxycortisol, etiocholanolone, pregnenolone, pregnanetriol, pregnanediol, and 5-pregnenediol) in human urine was developed and validated based on international guidelines. Steroids were enzymatically deconjugated and extracted by solid phase extraction before LC-MS/MS quantification in positive electrospray ionization mode.

Results

Excellent linearity with R2 > 0.99 and intra- and inter-day precisions of < 10.1 % were found. Relative matrix effects were between 96.4 % and 101.6 % and relative recovery was between 98.2 % and 115.0 %. Sufficient pre-freeze stability for all steroids in urine was found at 20-25 °C for seven days and at 4-6 °C for up to 28 days. Samples were stable during long-term storage at -20 °C and -80 °C for 6 months. Conclusions: A sensitive and robust LC-MS/MS method for the quantification of 11 urinary steroids was developed and validated according to international guidelines. Pre-analytical matrix stability was evaluated and the suitability of the method for the analysis of clinical samples and prospective validation studies was shown.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。