Association of leadless pacing with ventricular and valvular function

无导线起搏与心室和瓣膜功能的关系

阅读:1

Abstract

BACKGROUND: Traditional transvenous pacemakers are associated with worsening tricuspid valve function due to lead-related leaflet impingement, as well as ventricular dysfunction related to electromechanical dyssynchrony from chronic right ventricular (RV) pacing. The association of leadless pacing with ventricular and valvular function has not been well established. We aimed to assess the association of leadless pacemaker placement with changes in valvular regurgitation and ventricular function. METHODS AND RESULTS: Echocardiographic features before and after leadless pacemaker implant were analyzed in consecutive patients who received a leadless pacemaker with pre- and postprocedure echocardiography at Duke University Hospital between November 2014 and November 2019. Valvular regurgitation was graded ordinally from 0 (none) to 3 (severe). Among 54 patients, the mean age was mean age was 70.1 ± 14.3 years, 24 (44%) were women, and the most frequent primary pacing indication was complete heart block in 24 (44%). The median RV pacing burden was 45.4 (interquartile range [IQR] 3.5-97.0). On echocardiogram performed 8.9 months (IQR 4.5-14.5) after implant, there was no change in the average severity of tricuspid regurgitation (mean change 0.07 ± 1.15, p = .64) from pre-procedure echocardiogram. We observed a decrease in the average left ventricular ejection fraction (LVEF) (52.3 ± 9.3 to 47.9 ± 12.1, p = .0019) and tricuspid annular plane systolic excursion (TAPSE) (1.8 ± 0.6 to 1.6 ± 0.4, p = .0437). Thirteen patients (24%) had absolute drop in LVEF of ≥10%. CONCLUSION: We did not observe short term worsening valvular function in patients with leadless pacemakers. However, consistent with the pathophysiologic impact of RV pacing, leadless pacing was associated with a reduction in biventricular function.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。