Experimental Validation of Comprehensive Calculation for High-Resolution Linear MALDI-TOF Mass Spectrometry

高分辨率线性MALDI-TOF质谱综合计算的实验验证

阅读:6
作者:Yi-Hong Cai, Chia-Chen Wang, Chih-Hao Hsiao, Yi-Sheng Wang

Abstract

This work discusses the effectiveness of the previously developed comprehensive calculation model to optimize linear MALDI-TOF mass spectrometers. The model couples space- and velocity-focusing to precisely analyze the flight-time distribution of ions and predict optimal experimental parameters for the highest mass resolving power. Experimental validation was conducted using a laboratory-made instrument to analyze CsI3 and angiotensin I ions in low to medium m/z range. The results indicate that the predicted optimal extraction voltage and delay were reasonably accurate and effective. In the low m/z range, the peak width obtained using optimal parameters reached the sub nanosecond range, corresponding to a mass resolving power of 10 000-17 000, or 20 000-34 000 if shot-to-shot random fluctuations were minimized by the dynamic data correction method. The observed optimal mass resolving power in the current experiment is 4.8-7.8 times that of commercial instruments. Practical limitations resulting in the gap between the observed and theoretical ultimate mass resolving power are discussed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。