Detection of EGFR Mutations From Plasma of NSCLC Patients Using an Automatic Cartridge-Based PCR System

利用自动化卡式PCR系统检测非小细胞肺癌患者血浆中的EGFR突变

阅读:2

Abstract

The introduction of liquid biopsies for the detection of EGFR mutations in non-small cell lung cancer patients (NSCLC) has revolutionized the clinical care. However, liquid biopsies are technically challenging and require specifically trained personnel. To facilitate the implementation of liquid biopsies for the detection of EGFR mutations from plasma, we have assessed a fully automated cartridge-based qPCR test that allows the automatic detection of EGFR mutations directly from plasma. We have analyzed 54 NSCLC patients and compared the results of the cartridge-base device to an FDA-approved assay. Detection of EGFR mutations was comparable but slightly lower in the cartridge-based device for L858R mutations (14/15 detected, 93%) and exon 19 deletions (18/20 detected, 90%). Unfortunately, 8/54 (15%) tests failed but increasing the proteinase K volume helped to recover 3/4 (75%) unsuccessful samples. In summary, the fully automated cartridge-based device allowed the detection of EGFR mutations directly from plasma in NSCLC patients with promising accuracy. However, protocol adjustments are necessary to reduce a high test failure rate.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。