Short-Term Precision and Repeatability of Radiofrequency Echographic Multi Spectrometry (REMS) on Lumbar Spine and Proximal Femur: An In Vivo Study

射频回声多光谱法(REMS)在腰椎和近端股骨上的短期精确性和重复性:一项体内研究

阅读:3

Abstract

To determine the short-term intra-operator precision and inter-operator repeatability of radiofrequency echographic multi-spectrometry (REMS) at the lumbar spine (LS) and proximal femur (FEM). All patients underwent an ultrasound scan of the LS and FEM. Both precision and repeatability, expressed as root-mean-square coefficient of variation (RMS-CV) and least significant change (LSC) were obtained using data from two consecutive REMS acquisitions by the same operator or two different operators, respectively. The precision was also assessed in the cohort stratified according to BMI classification. The mean (±SD) age of our subjects was 48.9 ± 6.8 for LS and 48.3 ± 6.1 for FEM. Precision was assessed on 42 subjects at LS and 37 subjects on FEM. Mean (±SD) BMI was 24.71 ± 4.2 for LS and 25.0 ± 4.84 for FEM. Respectively, the intra-operator precision error (RMS-CV) and LSC resulted in 0.47% and 1.29% at the spine and 0.32% and 0.89% at the proximal femur evaluation. The inter-operator variability investigated at the LS yielded an RMS-CV error of 0.55% and LSC of 1.52%, whereas for the FEM, the RMS-CV was 0.51% and the LSC was 1.40%. Similar values were found when subjects were divided into BMI subgroups. REMS technique provides a precise estimation of the US-BMD independent of subjects' BMI differences.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。