Accuracy of ultrasound-guided versus palpation-based carpometacarpal joint injections: A randomized pilot study in cadavers

超声引导与触诊法腕掌关节注射的准确性比较:一项尸体随机对照试验

阅读:2

Abstract

OBJECTIVE: To determine the accuracy of ultrasound guidance compared to palpation in performing carpometacarpal joint injections in cadavers. DESIGN: In all, 36 carpometacarpal joints were randomized to either ultrasound-guided or palpation-based injections, with 1 cc of blue latex solution injected into each joint. The specimens were then dissected and the distribution of the latex was assessed by two independent, blinded raters. Injection accuracy was evaluated on a four-point quartile rating scale of 1-4, corresponding to the amount of the latex solution within the joint (1 = 0-25%, 2 = 26-50%, 3 = 51-75%, 4 = 76-100%). Inter-rater reliability was a secondary measure. RESULTS: The mean rating of accuracy was 2.1 for both palpation-based and ultrasound-guided injections. There was no statistically significant difference in accuracy between the two injectors. Chi-square analysis testing differences in accuracy for the two conditions was not statistically significant. The Cronbach's alpha for rater 2 was 0.74, which represents an acceptable level of reliability. A Friedman's Chi-square for the two raters was 2.3 (p = 0.13), indicating no significant difference between raters. CONCLUSION: Ultrasound guidance did not improve the accuracy of carpometacarpal joint injections in cadavers. However, the high inter-rater reliability attests to the value of the novel assessment scale.