Detecting failed elements on phased array ultrasound transducers using the Edinburgh Pipe Phantom

利用爱丁堡管道模型检测相控阵超声换能器上的故障元件

阅读:3

Abstract

AIMS: Imaging faults with ultrasound transducers are common. Failed elements on linear and curvilinear array transducers can usually be detected with a simple image uniformity or 'paperclip' test. However, this method is less effective for phased array transducers, commonly used in cardiac imaging. The aim of this study was to assess whether the presence of failed elements could be detected through measurement of the resolution integral (R) using the Edinburgh Pipe Phantom. METHODS: A 128-element paediatric phased array transducer was studied. Failed elements were simulated using layered polyvinyl chloride (PVC) tape as an attenuator and measurements of resolution integral were carried out for several widths of attenuator. RESULTS: All widths of attenuator greater than 0.5 mm resulted in a significant reduction in resolution integral and low contrast penetration measurements compared to baseline (p < 0.05). CONCLUSIONS: Measurements of resolution integral and low contrast penetration both have the potential to be used as straightforward and inexpensive tests to detect failed elements on phased array transducers. Particularly encouraging is the result for low contrast penetration as this is a quick and simple measurement to make and can be performed with many different test objects, thus enabling 'in-the-field' checks.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。