MRI-guided coupling for a focused ultrasound system using a top-to-bottom propagation

利用自上而下传播的MRI引导耦合技术实现聚焦超声系统

阅读:1

Abstract

BACKGROUND: A novel magnetic resonance imaging (MRI)-conditional coupling system was developed that accommodates a focused ultrasound (FUS) transducer. With this coupling system, the transducer can access targets from top to bottom. The intended clinical application is treatment of fibroids using FUS with the patient placed in supine position. METHODS: The coupling system was manufactured using a rapid prototyping device using acrylonitrile butadiene styrene (ABS) plastic. Coupling to a gel phantom was achieved using a water bag filled with degassed water. The FUS transducer was immersed in the water bag. RESULTS: The coupling system was successfully tested for MRI compatibility using fast-gradient pulse sequences in a gel phantom. The robotic system with its new coupling system was evaluated for its functionality for creating discrete and multiple (overlapping) lesions in the gel phantom. CONCLUSIONS: An MRI-conditional FUS coupling system integrated with an existing robotic system was developed that has the potential to create thermal lesions in targets using a top-to-bottom approach. This system has the potential to treat fibroid tumors with the patient lying in supine position.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。