Feasibility study for removing calcified material using a planar rectangular ultrasound transducer

利用平面矩形超声换能器去除钙化物质的可行性研究

阅读:2

Abstract

BACKGROUND: The aim of the proposed study was to conduct a feasibility study using a flat rectangular (3 mm × 10 mm) MRI compatible transducer operating at 5.3 MHz for destroying calcified material in an in vitro model. The proposed method can be used in the future for treating atherosclerosis plaques of the coronary, carotid or peripheral arteries. METHODS: The system was tested initially on calcium rods. Another test was performed in a hydroxyapatite-polylactide model. RESULTS: A parametric study was performed where the mass of calcified material removed was studied as a function of intensity, pulse repetition frequency (PRF), duty factor (DF) and presence of bubbles. CONCLUSIONS: The amount of calcified material removed is directly related to the intensity, PRF and DF. It was found that the presence of bubbles accelerates the removal of calcified material. In order to ensure that pure mechanical mode ultrasound was used, the protocols were designed so that the temperature does not exceed 1 °C.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。