Magnetic Resonance-guided High Intensity Focused Ultrasound in the presence of biopsy markers

磁共振引导高强度聚焦超声在活检标记物存在下的应用

阅读:1

Abstract

BACKGROUND: Magnetic Resonance guided High Intensity Focused ultrasound (MR-HIFU) offers precise non-invasive thermotherapy for clinical applications such as the treatment of breast lesions. However, patients with a biopsy marker are usually not eligible for MR-HIFU treatment. This study investigates the interaction of some MR-compatible markers with MR-HIFU thermotherapy. METHODS: The MR-HIFU compatibility of 14 markers (6 Gold Anchor and 4 Visicoil markers in gold, 1 Visicoil marker in brass, 3 BiomarC markers in carbon coated) were tested using the Sonalleve breast MR-HIFU platform at 1.5 T. The impact of these markers was assessed by counting the number of voxels with low signal intensity on MR thermal maps and by comparing temperature increases induced by the HIFU beam. RESULTS: Most markers were visible on thermal maps with an apparent size 4.2 ± 3.1 and 2 ± 1.8 times larger than their respective actual width and length. The volume of masked voxels was for most of the markers much larger than the actual volume of the marker (up to a factor 65.1). However, it represents only a small fraction of the 12 mm diameter targeted region (up to 8.8 voxels which represents 19% of this targeted region). Some differences in the maximal temperature increase were observed especially for BiomarC 1 × 3 and BiomarC 2 × 4 markers enhancing the heating. These differences were less pronounced at the edge of the targeted region. CONCLUSION: All markers had a minimal impact on the volume above the thermal dose threshold of 240 EM since the differences measured were smaller than the in-plane image resolution of 1.56 mm.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。