Classifying signals from a wearable accelerometer device to measure respiratory rate

对来自可穿戴加速度计设备的信号进行分类以测量呼吸频率

阅读:2

Abstract

BACKGROUND: Automatic measurement of respiratory rate in general hospital patients is difficult. Patient movement degrades the signal and variation of the breathing cycle means that accurate observation for ≥60 s is needed for adequate precision. METHODS: We studied acutely ill patients recently admitted to a teaching hospital. Breath duration was measured from a triaxial accelerometer attached to the chest wall and compared with a signal from a nasal cannula. We randomly divided the patient records into a training (n=54) and a test set (n=7). We used machine learning to train a neural network to select reliable signals, automatically identifying signal features associated with accurate measurement of respiratory rate. We used the test records to assess the accuracy of the device, indicated by the median absolute difference between respiratory rates, provided by the accelerometer and by the nasal cannula. RESULTS: In the test set of patients, machine classification of the respiratory signal reduced the median absolute difference (interquartile range) from 1.25 (0.56-2.18) to 0.48 (0.30-0.78) breaths per min. 50% of the recording periods were rejected as unreliable and in one patient, only 10% of the signal time was classified as reliable. However, even only 10% of observation time would allow accurate measurement for 6 min in an hour of recording, giving greater reliability than nurse charting, which is based on much less observation time. CONCLUSION: Signals from a body-mounted accelerometer yield accurate measures of respiratory rate, which could improve automatic illness scoring in adult hospital patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。