Fast quantitative three-dimensional ultrashort echo time (UTE) Cones magnetic resonance imaging of major tissues in the knee joint using extended sprial sampling

利用扩展螺旋采样技术,对膝关节主要组织进行快速定量三维超短回波时间(UTE)锥形磁共振成像。

阅读:2

Abstract

The purpose of this study is to investigate the effect of extending the spiral sampling window on quantitative 3D ultrashort echo time (UTE) Cones imaging of major knee joint tissues including articular cartilage, menisci, tendons and ligaments at 3 T. Nine cadaveric human whole knee specimens were imaged on a 3 T clinical MRI scanner. A series of quantitative 3D UTE Cones imaging biomarkers including T(2) *, T(1) , adiabatic T(1ρ) , magnetization transfer ratio (MTR) and macromolecular fraction (MMF) were estimated using spiral sampling trajectories with various durations. Errors in UTE MRI biomarkers as a function of sampling time were evaluated using a nonstretched spiral trajectory as a reference standard. No significant differences were observed by increasing the spiral sampling window from 1116 to 2232 μs in the calculated T(2) *, T(1) , adiabatic T(1ρ) , MTR and MMF, as all P-values were over .05 as assessed by ANOVA with two-sided Dunnett's test. Although extending the sampling window results in signal loss for short T(2) components, there was limited effect on the calculated quantitative biomarkers, with error percentages typically smaller than 5% in all the evaluated tissues. The total scan time can be reduced by up to 54% with quantification errors of less than 5% in any evaluated major tissue in the knee joint, suggesting that 3D UTE Cones MRI techniques can be greatly accelerated by using a longer spiral sampling window without causing additional quantitative bias.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。