Screen-printed ion-selective electrodes were designed and characterized for the assessment of cyromazine (CYR) pesticide. A novel approach is to design tailor-made specific recognition sites in polymeric membranes using molecularly imprinted polymers for cyromazine (CR) determination (sensor I). Another sensor (sensor II) is the plasticized PVC membrane incorporating cyromazine/tetraphenyl borate ion association complex. The charge-transfer resistance and water layer reached its minimal by incorporating Polyaniline (PANI) solid-contact ISE. The designed electrodes demonstrated Nernstain response over a linear range 1.0 × 10(-2)-5.2 × 10(-6) and 1.0 × 10(-2)-5.7 × 10(-5) M with a detection limit 2.2 × 10(-6) and 8.1 × 10(-6) M for sensors I and II, respectively. The obtained slopes were 28.1 ± 2.1 (r(2) = 0.9999) and 36.4 ± 1.6 (r(2) = 0.9991) mV/decade, respectively. The results showed that the proposed electrodes have a fast and stable response, good reproducibility, and applicability for direct measurement of CYR content in commercial pesticide preparations and soil samples sprayed with CYR pesticide. The results obtained from the proposed method are fairly in accordance with those using the standard official method.