Evaluation of Cleaning Methods for Lithium Disilicate Ceramic Post Try-In Paste Application: An SEM Analysis

二硅酸锂陶瓷桩试戴膏应用清洁方法的评价:扫描电镜分析

阅读:2

Abstract

This in vitro study assessed the efficacy of three cleaning methods on lithium disilicate ceramic after the application of different try-in pastes through SEM analysis. Ten rectangular specimens of IPS e.max CAD were prepared using a diamond disc, crystallized, etched with 5% hydrofluoric acid, and subjected to three try-in pastes-Calibra ©, Variolink (V), RelyX Veneer(®)-and three cleaning techniques-air-water spray (RD), ultrasonic bath in distilled water for five minutes (ULT/W), and ultrasonic bath in distilled alcohol for five minutes (ULT/A). A control specimen was also included. After one-minute paste application and subsequent cleaning method application, SEM evaluation was conducted. The results indicate that RD was as effective as CTRL in removing remnants from R-RD, V-ULT/W and V-ULT/A samples, but ineffective for all Calibra paste-contaminated specimens. In conclusion, the optimal removal of try-in paste residues from lithium disilicate restorations is paste-dependent; however, ultrasonic baths with distilled water or alcohol proved effective for most pastes tested.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。