Conclusion
These complementary new approaches provide a simple means to obtain information to assist diagnosis at the POC. The first provides antibiotic sensitivity following 10 hours incubation, whereas the second optical approach (QuickCheck), provides instant accurate total leukocyte count.
Methods
One approach uses a tetrazolium-based chemical reporting system, primarily focused on detecting bacterial contamination and associated vancomycin-sensitivity. The second approach uses a novel forward light-scatter device (QuickCheck) to provide an instant quantitative cell count directly from PD patient effluent.
Results
The tetrazolium approach detected and correctly distinguished laboratory isolates, taking 10 hours to provide non-quantitative results. We compared the technical performance of the light scatter leukocyte counting approach with spectrophotometry, hemocytometer counting and flow cytometry (Sysmex) using patient effluent samples. QuickCheck had high accuracy (94%) and was the most precise (coefficient of variation <4%), showing minimal bias, overall performing similarly to flow cytometry.