Establishment and Evaluation of a Novel Method Based on Loop-Mediated Isothermal Amplification for the Rapid Diagnosis of Thalassemia Genes

基于环介导等温扩增的地中海贫血基因快速诊断新方法的建立及评价

阅读:6
作者:Wei-Hua Wang #, Min Lin #, Hai-Liang Li, Jun-Yun Huang, Jiang-Tao Chen, Xian-Song Fang, Dong-Mei Huang, Xu-Xiang Xi, Qing-Fei Zhao, Fang-Li Song, Shao Huang, Tian-Yu Zhong

Conclusion

The novel LAMP assay can be completed within 60 min using a heating block or a water bath, and the result can be read visually based on color change to detect thalassemia. The LAMP assay fulfills the requirements of field application and resource-limited areas, especially those with primary hospitals and rural areas.

Methods

Primer sequences were designed using Primer Explorer V4 software. Blood samples (5 mL) were collected from all participants in EDTA. DNA was extracted using Chelex 100 and was subjected to LAMP. LAMP products were detected by fluorescence development in ultraviolet light.

Purpose

Currently, thalassemia is commonly detected using gap-polymerase chain reaction (PCR) and deoxyribonucleic acid (DNA) reverse dot blot, which have high requirements of space, instruments, and personnel. Therefore, it is necessary to develop a new method for thalassemia detection with high sensitivity, low cost, and simple and fast operation. In this study, we aimed to design and evaluate a new method for detecting three α-thalassemia genes including -Southeast Asian (SEA), -α3.7, and -α4.2 and five β-thalassemia genes including 654M, 41/42M, -28M, 17M, and 27/28M based on loop-mediated isothermal amplification (LAMP).

Results

We found that LAMP assays for positive samples of thalassemia reached a plateau before 60 minutes, whereas the negative control samples entered the plateau after 70 minutes or showed no amplification. The concentration range of positive reactions was between 20-60 pg/μL and 20-60 ng/μL. Additionally, there were no cross-reactivities among 8 thalassemia subtypes. For clinical samples, the positive sample tube showed strong green fluorescence, whereas the negative tube showed light green fluorescence. According to these results, the LAMP method has high sensitivity for detecting thalassemia (252/254). However, 43 false-positive results were obtained in the LAMP test. The LAMP assay was also of low cost and with simple and fast operation.