Validation of Decisions for Percutaneous Coronary Intervention Using Smartphone-Based Electrocardiogram Device Spandan: A Cross-Sectional Observational Study

利用智能手机心电图设备Spandan验证经皮冠状动脉介入治疗决策:一项横断面观察研究

阅读:3

Abstract

BACKGROUND: India bears a high burden of acute coronary syndrome, with younger patients and a high prevalence of ST-elevation myocardial infarction (STEMI). Spandan is, therefore, an attractive smartphone-based electrocardiogram (ECG) device that could allow for potentially early diagnosis as well as enabling timely intervention which may even save lives in resource-poor settings. The study aimed to assess the performance and diagnostic capability of the Spandan smartphone-based ECG device in decision-making for percutaneous coronary intervention (PCI) by analyzing the initial ST-segment elevation, which was compared to a 12-lead ECG as the gold standard (BPL Cardiart ECG Machine). METHODS: This was an observational cross-sectional study involving 184 eligible participants with chest pain presenting to the local hospital, in Meerut, Uttar Pradesh, India. The study was conducted for the evaluation of the diagnostic appropriateness of the Spandan ECG device for the detection of ST elevation as compared to standard 12-lead ECGs so that the cardiologists could be more easily guided in their decisions relative to PCI. Patients with the onset of chest pain within or after 120 h and ST elevation above 1 mm in two or more leads were enrolled and patients with dementia, bundle branch block, cardiogenic shock, and ECG artifacts were excluded. The analysis included calculating response characteristics and estimating correlation coefficients and confusion matrix to compare both appraisal methods. RESULTS: The Spandan device performed with good agreement with the gold standard ECG, particularly in the leads II, III, and AVF, with Pearson correlation coefficients close to 1. The ST elevation in the Spandan device showed no statistical difference compared to the 12-lead ECG. The device exhibited a sensitivity of 94% and a positive predictive value of 94% for ST-elevation detection, thus having supportive evidence for possible usefulness for decision-making in PCI. CONCLUSIONS: ECG findings, such as that of the smartphone-based device (Spandan Pro ECG, a single channel autoswitched ECG machine), demonstrated comparable accuracy with the gold standard 12-lead ECG for the diagnosis of ST elevation and helped in making clinical decisions in patients requiring PCI, especially in resource-limited settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。