Abstract
A new hydrogen peroxide (H(2)O(2)) sensor is fabricated based on a multiwalled carbon nanotube-modified glassy carbon electrode (MWCNT-GCE) and reactive blue 19 (RB). The charge transfer coefficient, α, and the charge transfer rate constant, k(s), of RB adsorbed on MWCNT-GCE were calculated and found to be 0.44 ± 0.01 Hz and 1.9 ± 0.05 Hz, respectively. The catalysis of the electroreduction of H(2)O(2) by RB-MWCNT-GCE is described. The RB-MWCNT-GCE shows a dramatic increase in the peak current and a decrease in the overvoltage of H(2)O(2) electroreduction in comparison with that seen at an RB modified GCE, MWCNT modified GCE, and activated GCE. The kinetic parameters such as α and the heterogeneous rate constant, k', for the reduction of H(2)O(2) at RB-MWCNT-GCE surface were determined using cyclic voltammetry. The detection limit of 0.27μM and three linear calibration ranges were obtained for H(2)O(2) determination at the RB-MWCNT-GCE surface using an amperometry method. In addition, using the newly developed sensor, H(2)O(2) was determined in real samples with satisfactory results.