Background
Patients with somatostatin receptor-expressing neuroendocrine tumours can be treated with intravenously administered 177Lu-octreotate. Few patients are cured with the present protocol due to the current dose limitation of normal organs at risk, such as the kidneys. By locally administering 177Lu-octreotate to the liver for the
Conclusions
Acceptable results were obtained for both detectors, although the semiconductor detector proved more advantageous for our purpose. The measurements demonstrated factors that must be corrected for, such as organ thickness or dead-time effects. Altogether, intraoperative gamma detector measurements could be used to determine 177Lu activity concentration in vivo.
Results
Measurements were made using different sources containing 177Lu. Response linearity, sensitivity, spatial resolution and its depth dependence, organ thickness dependence of the measured count rate and tumour detectability were assessed for two intraoperative gamma detectors. The two detectors (a scintillation and a semiconductor detector) showed differences in technical performance. For example, the sensitivity was higher for the scintillation detector, while the spatial resolution was better for the semiconductor detector. Regarding organ thickness dependence and tumour detectability, similar results were obtained for both detectors, and even relatively small simulated tumours of low tumour-to-background activity concentration ratios could be detected. Conclusions: Acceptable results were obtained for both detectors, although the semiconductor detector proved more advantageous for our purpose. The measurements demonstrated factors that must be corrected for, such as organ thickness or dead-time effects. Altogether, intraoperative gamma detector measurements could be used to determine 177Lu activity concentration in vivo.
