Multicenter closed-loop insulin delivery study points to challenges for keeping blood glucose in a safe range by a control algorithm in adults and adolescents with type 1 diabetes from various sites

多中心闭环胰岛素输送研究指出,通过控制算法将来自不同地点的 1 型糖尿病成人和青少年的血糖保持在安全范围内存在挑战

阅读:5
作者:Howard Zisser, Eric Renard, Boris Kovatchev, Claudio Cobelli, Angelo Avogaro, Revital Nimri, Lalo Magni, Bruce A Buckingham, H Peter Chase, Francis J Doyle 3rd, John Lum, Peter Calhoun, Craig Kollman, Eyal Dassau, Anne Farret, Jerome Place, Marc Breton, Stacey M Anderson, Chiara Dalla Man, Simone De

Background

The Control to Range Study was a multinational artificial pancreas study designed to assess the time spent in the hypo- and hyperglycemic ranges in adults and adolescents with type 1 diabetes while under closed-loop control. The controller attempted to keep the glucose ranges between 70 and 180 mg/dL. A set of prespecified metrics was used to measure safety. Research design and

Conclusions

The control-to-range system performed as expected: faring better overnight than during the day and performing with variability between patients even after individualization based on patients' prior settings. The system had difficulty preventing postmeal excursions above target range.

Methods

We studied 53 individuals for approximately 22 h each during clinical research center admissions. Plasma glucose level was measured every 15-30 min (YSI clinical laboratory analyzer instrument [YSI, Inc., Yellow Springs, OH]). During the admission, subjects received three mixed meals (1 g of carbohydrate/kg of body weight; 100 g maximum) with meal announcement and automated insulin dosing by the controller.

Results

For adults, the mean of subjects' mean glucose levels was 159 mg/dL, and mean percentage of values 71-180 mg/dL was 66% overall (59% daytime and 82% overnight). For adolescents, the mean of subjects' mean glucose levels was 166 mg/dL, and mean percentage of values in range was 62% overall (53% daytime and 82% overnight). Whereas prespecified criteria for safety were satisfied by both groups, they were met at the individual level in adults only for combined daytime/nighttime and for isolated nighttime. Two adults and six adolescents failed to meet the daytime criterion, largely because of postmeal hyperglycemia, and another adolescent failed to meet the nighttime criterion. Conclusions: The control-to-range system performed as expected: faring better overnight than during the day and performing with variability between patients even after individualization based on patients' prior settings. The system had difficulty preventing postmeal excursions above target range.

Trial registration

ClinicalTrials.gov NCT01271023.