Comparison of distortion-product otoacoustic emission growth rates and slopes of forward-masked psychometric functions

畸变产物耳声发射增长率与前向掩蔽心理测量函数斜率的比较

阅读:2

Abstract

Slopes of forward-masked psychometric functions (FM PFs) were compared with distortion-product otoacoustic emission (DPOAE) input/output (I/O) parameters at 1 and 6 kHz to test the hypothesis that these measures provide similar estimates of cochlear compression. Implicit in this hypothesis is the assumption that both DPOAE I/O and FM PF slopes are functionally related to basilar-membrane (BM) response growth. FM PF-slope decreased with signal level, but this effect was reduced or reversed with increasing hearing loss; there was a trend of decreasing psychometric function (PF) slope with increasing frequency, consistent with greater compression at higher frequencies. DPOAE I/O functions at 6 kHz exhibited an increase in the breakpoint of a two-segment slope as a function of hearing loss with a concomitant decrease in the level of the distortion product (L(d)). Results of the comparison between FM PF and DPOAE I/O parameters revealed only a weak correlation, suggesting that one or both of these measures may provide unreliable information about BM compression.