Clinical Management and Pump Parameter Adjustment of the Control-IQ Closed-Loop Control System: Results from a 6-Month, Multicenter, Randomized Clinical Trial

Control-IQ 闭环控制系统的临床管理和泵参数调整:6 个月、多中心、随机临床试验的结果

阅读:6
作者:Grenye O'Malley, Laurel H Messer, Carol J Levy, Jordan E Pinsker, Gregory P Forlenza, Elvira Isganaitis, Yogish C Kudva, Laya Ekhlaspour, Dan Raghinaru, John Lum, Sue A Brown; iDCL Trial Research Group

Background

Data are limited on the need for and benefits of pump setting optimization with automated insulin delivery. We examined clinical management of a closed-loop control (CLC) system and its relationship to glycemic outcomes. Materials and

Conclusions

No difference in TIR was detected with parameter changes overall, but they may have an effect in higher HbA1c subgroups or following user-directed boluses, suggesting that changes may matter more in suboptimal control or during discrete periods of the day. Clinical Trials Registration number: NCT03563313.

Methods

We analyzed personal parameter adjustments in 168 participants in a 6-month multicenter trial of CLC with Control-IQ versus sensor-augmented pump (SAP) therapy. Preset parameters (BR = basal rates, CF = correction factors, CR = carbohydrate ratios) were optimized at randomization, 2 and 13 weeks, for safety issues, participant concerns, or initiation by participants' usual diabetes care team. Time in range (TIR 70-180 mg/dL) was compared in the week before and after parameter changes.

Results

In 607 encounters for parameter changes, there were fewer adjustments for CLC than SAP (3.4 vs. 4.1/participant). Adjustments involved BR (CLC 69%, SAP 80%), CR (CLC 68%, SAP 50%), CF (CLC 44%, SAP 41%), and overnight parameters (CLC 62%, SAP 75%). TIR before and after adjustments was 71.2% and 71.3% for CLC and 61.0% and 62.9% for SAP. The highest baseline HbA1c CLC subgroup had the largest TIR improvement (51.2% vs. 57.7%). When a CR was made more aggressive in the CLC group, postprandial time >180 mg/dL was 43.1% before the change and 36.0% after the change. The median postprandial time <70 mg/dL before making CR less aggressive was 1.8%, and after the change was 0.7%. Conclusions: No difference in TIR was detected with parameter changes overall, but they may have an effect in higher HbA1c subgroups or following user-directed boluses, suggesting that changes may matter more in suboptimal control or during discrete periods of the day. Clinical Trials Registration number: NCT03563313.