Biosensor-based fragment screening using FastStep injections

使用 FastStep 注射进行基于生物传感器的片段筛选

阅读:8
作者:Rebecca L Rich, John G Quinn, Tom Morton, J David Stepp, David G Myszka

Abstract

We have developed a novel analyte injection method for the SensíQ Pioneer surface plasmon resonance-based biosensor referred to as "FastStep." By merging buffer and sample streams immediately prior to the reaction flow cells, the instrument is capable of automatically generating a two- or threefold dilution series (of seven or five concentrations, respectively) from a single analyte sample. Using sucrose injections, we demonstrate that the production of each concentration within the step gradient is highly reproducible. For kinetic studies, we developed analysis software that utilizes the sucrose responses to automatically define the concentration of analyte at any point during the association phase. To validate this new approach, we compared the results of standard and FastStep injections for ADP binding to a target kinase and a panel of compounds binding to carbonic anhydrase II. Finally, we illustrate how FastStep can be used in a primary screening mode to obtain a full concentration series of each compound in a fragment library.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。