Design and Characterization of an Osmotic Sensor for the Detection of Events Associated with Dehydration and Overhydration

用于检测脱水和过度补水相关事件的渗透传感器的设计和特性

阅读:5
作者:Luís André Fernandes, Philipp Häfliger, Mehdi Azadmehr, Erik Johannessen

Abstract

The level of hydration in the human body is carefully adjusted to control the electrolyte balance that governs the biochemical processes that sustain life. An electrolyte deficiency caused by de- or overhydration will not only limit human performance, but can also lead to serious health problems and death if left untreated. Because humans can withstand a change in hydration of only [Formula: see text], frequent monitoring should be performed in risk groups. This paper presents an osmotic hydration sensor that can record the level of hydration as a function of osmotic pressure in phosphate buffered saline or sodium-chloride solutions that simulate the interstitial fluid in the body. The osmotic pressure is recorded with the aid of an ion-exchange membrane that facilitates the migration of water and cations, in favor of reverse osmosis or gas separation membranes. The hydration sensor is designed to be coupled to an inductively powered readout circuit designed for integration in a micro-implant that has previously been shown to consume only 76 [Formula: see text] of power. The dynamic range spans a state of serious overhydration (220 [Formula: see text]) to a serious state of dehydration (340 [Formula: see text]) with a response time of [Formula: see text] (for a variation of hydration of 20%).