Calorimetric Observation of Single He2∗ Excimers in a 100-mK He Bath

在 100 mK 氦浴中对单个 He2* 准分子进行量热观测

阅读:1

Abstract

We report the first calorimetric detection of individual He2∗ excimers within a bath of superfluid 4He . The detector used in this work is a single superconducting titanium transition edge sensor (TES) with an energy resolution of  ∼ 1 eV , immersed directly in the helium bath. He2∗ excimers are produced in the surrounding bath using an external gamma-ray source. These excimers exist either as short-lived singlet or long-lived triplet states. We demonstrate detection (and discrimination) of both states: In the singlet case the calorimeter records the absorption of a prompt  ≈ 15 eV photon, and in the triplet case the calorimeter records a direct interaction of the molecule with the TES surface, which deposits a distinct fraction of the  ≈ 15 eV , released upon decay, into the surface. We also briefly discuss the detector fabrication and characterization.