Hemodynamic performance limits of the neonatal Double-Lumen cannula

新生儿双腔插管的血流动力学性能极限

阅读:1

Abstract

Venovenous extracorporeal membrane oxygenation (VV-ECMO) is the preferred surgical intervention for patients suffering from severe cardiorespiratory failure, also encountered in SARS-Cov-2 management. The key component of VV-ECMO is the double-lumen cannula (DLC) that enables single-site access. The biofluid dynamics of this compact device is particularly challenging for neonatal patients due to high Reynolds numbers, tricuspid valve location and right-atrium hemodynamics. In this paper we present detailed findings of our comparative analysis of the right-atrial hemodynamics and salient design features of the 13Fr Avalon Elite DLC (as the clinically preferred neonatal cannula) with the alternate Origen DLC design, using experimentally validated computational fluid dynamics. Highly accurate 3D-reconstructions of both devices were obtained through an integrated optical coherence tomography and micro-CT imaging approach. Both cannula configurations displayed complex flow structures inside the atrium, superimposed over predominant recirculation regimes. We found that the Avalon DLC performed significantly better than the Origen alternative, by capturing 80% and 94% of venous blood from the inferior and superior vena cavae, respectively and infusing the oxygenated blood with an efficiency of more than 85%. The micro-scale geometric design features of the Avalon DLC that are associated with superior hemodynamics were investigated through 14 parametric cannula configurations. These simulations showed that the strategic placement of drainage holes, the smooth infusion blood stream diverter and efficient distribution of the venous blood capturing area between the vena cavae are associated with robust blood flow performance. Nevertheless, our parametric results indicate that there is still room for further device optimization beyond the performance measurements for both Avalon and Origen DLC in this study. In particular, the performance envelope of malpositioned cannula and off-design conditions require additional blood flow simulations for analysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。