Reliable calibration by nonlinear standard addition method in the presence of additive interference effects

在存在加性干扰效应的情况下,采用非线性标准加入法进行可靠校准

阅读:1

Abstract

ABSTRACT: The possibility of adapting the Standard Addition Method (SAM) to calibration in very difficult analytical conditions, namely when there is a need to determine an analyte with the use of nonlinear calibration graph and in the presence of matrix components causing additive interference effect, is investigated. To this aim the SAM in the common version and the Chemical H-point Standard Addition Method (C-HPSAM) realized by the flow injection technique were applied. Specifically, a flow manifold was used for construction of a set of nonlinear calibration graphs in different chemical conditions. As the graphs were intersected indicating both the additive interference effect and the analytical result free of this effect, the analyte concentration in the sample was able to be obtained with improved accuracy. The applicability of this approach was verified on the example of spectrophotometric determination of paracetamol in pharmaceuticals and of total acidity in wines. The C-HPSAM method enabled complete compensation of the additive effect and obtaining analytical results at a relative error not exceeding 6.0%.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。