Hyperspectral imaging of 4T1 mammary carcinomas grown in dorsal skinfold window chambers: spectral renormalization and fluorescence modeling

背部皮褶窗室中生长的 4T1 乳腺癌的高光谱成像:光谱重正化和荧光建模

阅读:23
作者:Tadej Tomanic, Tim Bozic, Bostjan Markelc, Jost Stergar, Gregor Sersa, Matija Milanic

Aim

We highlight the significance of spectral renormalization in HSI of DSWC models and demonstrate the benefit of incorporating enhanced green fluorescent protein (EGFP) excitation and emission in the skin tissue model for tumors expressing genes to produce EGFP. Approach: We employed an HSI system for intravital imaging of mice with 4T1 mammary carcinoma in a DSWC over 14 days. We performed spectral renormalization of hyperspectral images based on the measured reflectance spectra of glass coverslips and utilized an inverse adding-doubling (IAD) algorithm with a two-layer murine skin model, to extract tissue parameters, such as total hemoglobin concentration and tissue oxygenation ( StO2<math> <mrow><msub><mi>StO</mi> <mn>2</mn></msub> </mrow> </math> ). The model was upgraded to consider EGFP fluorescence excitation and emission. Moreover, we conducted additional experiments involving tissue phantoms, human forearm skin imaging, and numerical simulations.

Conclusions

The results contribute to improving intravital monitoring of DWSC models using HSI and pave the way for more accurate and precise quantitative imaging.

Results

Hyperspectral image renormalization and the addition of EGFP fluorescence in the murine skin model reduced the mean absolute percentage errors (MAPEs) of fitted and measured spectra by up to 10% in tissue phantoms, 0.55% to 1.5% in the human forearm experiment and numerical simulations, and up to 0.7% in 4T1 tumors. Similarly, the MAPEs for tissue parameters extracted by IAD were reduced by up to 3% in human forearms and numerical simulations. For some parameters, statistically significant differences ( p<0.05<math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ) were observed in 4T1 tumors. Ultimately, we have shown that fluorescence emission could be helpful for 4T1 tumor segmentation. Conclusions: The results contribute to improving intravital monitoring of DWSC models using HSI and pave the way for more accurate and precise quantitative imaging.

Significance

Hyperspectral imaging (HSI) of murine tumor models grown in dorsal skinfold window chambers (DSWCs) offers invaluable insight into the tumor microenvironment. However, light loss in a glass coverslip is often overlooked, and particular tissue characteristics are improperly modeled, leading to errors in tissue properties extracted from hyperspectral images. Aim: We highlight the significance of spectral renormalization in HSI of DSWC models and demonstrate the benefit of incorporating enhanced green fluorescent protein (EGFP) excitation and emission in the skin tissue model for tumors expressing genes to produce EGFP. Approach: We employed an HSI system for intravital imaging of mice with 4T1 mammary carcinoma in a DSWC over 14 days. We performed spectral renormalization of hyperspectral images based on the measured reflectance spectra of glass coverslips and utilized an inverse adding-doubling (IAD) algorithm with a two-layer murine skin model, to extract tissue parameters, such as total hemoglobin concentration and tissue oxygenation ( StO2<math> <mrow><msub><mi>StO</mi> <mn>2</mn></msub> </mrow> </math> ). The model was upgraded to consider EGFP fluorescence excitation and emission. Moreover, we conducted additional experiments involving tissue phantoms, human forearm skin imaging, and numerical simulations. Results: Hyperspectral image renormalization and the addition of EGFP fluorescence in the murine skin model reduced the mean absolute percentage errors (MAPEs) of fitted and measured spectra by up to 10% in tissue phantoms, 0.55% to 1.5% in the human forearm experiment and numerical simulations, and up to 0.7% in 4T1 tumors. Similarly, the MAPEs for tissue parameters extracted by IAD were reduced by up to 3% in human forearms and numerical simulations. For some parameters, statistically significant differences ( p<0.05<math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ) were observed in 4T1 tumors. Ultimately, we have shown that fluorescence emission could be helpful for 4T1 tumor segmentation. Conclusions: The results contribute to improving intravital monitoring of DWSC models using HSI and pave the way for more accurate and precise quantitative imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。