Label-free Growth Receptor-2 Detection and Dissociation Constant Assessment in Diluted Human Serum Using a Longitudinal Extension Mode of a Piezoelectric Microcantilever Sensor

使用压电微悬臂梁传感器的纵向延伸模式进行无标记生长受体-2 检测和稀释人血清中的解离常数评估

阅读:13
作者:Joseph A Capobianco, Wan Y Shih, Gregory P Adams, Wei-Heng Shih

Abstract

We have investigated real-time, label-free, in-situ detection of human epidermal growth factor receptor 2 (Her2) in diluted serum using the first longitudinal extension mode of a lead zirconate-lead titanate (PZT)/glass piezoelectric microcantilever sensor (PEMS) with H3 single-chain variable fragment (scFv) immobilized on the 3-mercaptopropyltrimethoxysilane (MPS) insulation layer of the PEMS surface. We showed that with the longitudinal extension mode, the PZT/glass PEMS consisting of a 1 mm long and 127 μm thick PZT layer bonded with a 75 μm thick glass layer with a 1.8 mm long glass tip could detect Her2 at a concentration of 6-60 ng/ml (or 0.06-0.6 nM) in diluted human serum, about 100 times lower than the concentration limit obtained using the lower-frequency flexural mode of a similar PZT/glass PEMS. We further showed that with the longitudinal mode, the PZT/glass PEMS determined the equilibrium H3-Her2 dissociation constant K(d) to be 3.3±0.3 × 10(-8) M consistent with the value, 3.2±0.28 ×10(-8) M deduced by the surface plasmon resonance method (BIAcore).