Evaluation of Higher-Order Aberrations After the Smooth Incision Lenticular Keratomileusis (SILK(TM)) Procedure Using the ELITA(TM) Femtosecond Platform for Correction of Myopic and Astigmatic Refractive Errors

使用ELITA™飞秒平台对光滑切口角膜磨镶术(SILK™)矫正近视和散光屈光不正后高阶像差的评估

阅读:2

Abstract

PURPOSE: To evaluate the changes of higher-order wavefront aberrations following the Smooth Incision Lenticular Keratomileusis (SILK(TM)) procedure for correction of myopic refractive errors with and without astigmatism, using the ELITA(TM) Femtosecond Platform. METHODS: This prospective study included 24 eyes that underwent SILK procedure using one ELITA femtosecond laser system for the correction of myopic refractive errors with and without astigmatism. Preoperative and postoperative 1-day, 1-week, 1-month, 3-month, and 9-month eye exams were measured with a commercial wavefront aberrometer (iDESIGN (®) Refractive Studio, Johnson & Johnson Surgical Vision, Inc). Wavefront aberrations up to the 6th order Zernike coefficients, including coma Z(3, -1) and Z(3, 1), spherical aberration Z(4, 1), and the wavefront error of all higher-order aberrations (HOAs RMS), were evaluated across a 6 mm pupil. RESULTS: The mean manifest refractive spherical equivalent changed from the preoperative refractions -3.82 ± 1.26 D (range -6.00 to -2.25 D) to the postoperative refractions -0.20 ± 0.15 D (range -0.50 to 0.00 D) at the 9-month follow-up. Compared to baseline preoperative HOAs, the mean postoperative HOAs were significantly increased at the 1-day follow-up. On average, at the 9-month postoperative assessment the vertical coma Z(3, -1) was -0.054 ±0.186 µm, horizontal coma Z(3, 1) was 0.016 ± 0.124 µm, spherical aberration Z(4, 0) was 0.046 ± 0.163 µm, and HOAs RMS was 0.363 ± 0.115 µm across a 6 mm pupil. There is no significant difference in the mean HOAs starting at 1-week follow-up for the horizontal coma (P = 0.346) and spherical aberration (P = 0.095). CONCLUSIONS: The visual outcomes demonstrated that the SILK procedure for refractive lenticule extraction using ELITA femtosecond laser system is effective and predictable for the correction of myopic refractive errors with and without astigmatism. The ELITA femtosecond laser system induced minimal HOAs in surgical eyes following the SILK procedures. These results demonstrate fast corneal recovery starting at 1-week follow-up, and spherical aberration was not induced.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。